Setup examples, Setting the pseudo log range (logtype2) – Yokogawa Value Series FX1000 User Manual
Page 107
3-34
IM 04L21B01-01EN
Setup Examples
LogType2 [pseudo log] and LogType1 [nonlinear log] setup examples are shown below.
• Setting the Pseudo Log Range (LogType2)
This setting is for logging the pseudo log output (see the table below) of a vacuum gauge.
Pressure Reading [Pa] Output Voltage [V] Pressure Reading [Pa] Output Voltage [V]
1.30x10
-7
0.13
5.00x10
-4
3.50
5.00x10
-7
0.50
1.00x10
-3
4.10
1.00x10
-6
1.10
5.00x10
-3
4.50
5.00x10
-6
1.50
1.00x10
-2
5.10
1.00x10
-5
2.10
5.00x10
-2
5.50
5.00x10
-5
2.50
1.00x10
-1
6.10
1.00x10
-4
3.10
9.90x10
-1
6.99
1
Press
MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to
basic setting mode), and select the
Environment tab > LOG input
Set
LogType2 under Input to Pseudo.
2
Press
MENU (to switch to setting mode), and select the Menu tab > Meas
channel > Range, Alarm.
Set the range as shown below.
Mode: LogType2
Range: 20V
Span: Lower 0.000, Upper 7.000
Scale: Lower 1.00E-7, Upper 1.00E0
Display Example
3.13 Using the Log Scale to Perform Measurements (/LG1 Option)