2.3 Setting the Time Correction Operation during

Memory Sampling

This function gradually corrects the time when the time is changed while Memory
Sampling is in progress.
For a description of the time correction operation, see section 1.10.

Setup Screen

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic

setting mode), and select the

Menu tab > Time Settings.

Setup Items

• Time settings > Time deviation limit

Time deviation limit

Current time

Specified time

Only when the time deviation between the FX’s time and the specified time is within

± (the value that you specify here), the FX’s time gradually corrected. Otherwise, the
clock is corrected immediately.

Settings

Description

10 s to 5 min

The time deviation limit.

Off

Disables the function that gradually corrects the time.

Example: If Time deviation limit is set to 10s and the time on the FX is 10:21:15,

the time on the FX will be gradually corrected to the specified time if the
specified time is between 10:21:05 and 10:21:25.