Extended input (/n3f), Remote control (/r1) – Yokogawa Value Series FX1000 User Manual
Page 277
13-12
IM 04L21B01-01EN
Extended Input (/N3F)
Item
Specifications
Measurement/display accuracy
Under standard operating conditions
Input Type
Measuring Range
Measurement Accuracy
Max.
Resolution
Kp vs Au7Fe
0.0 to 300.0 K
0 to 20 K
Within ±4.5 K
0.1 K
20 to 300 K
Within ±2.5 K
PLATINEL
0.0 to 1400.0°C
±(0.25% of rdg + 2.3°C)
0.1°C
PR40-20
0.0 to 1900.0°C
0 to 450°C
Accuracy not guaranteed
450 to 750°C
±(0.9% of rdg + 3.2°C)
750 to 1100°C
±(0.9% of rdg + 1.3°C)
1100 to 1900°C
±(0.9% of rdg + 0.4°C)
Thermocouple NiNiMo
0.0 to 1310.0°C
±(0.25% of rdg + 0.7°C)
W/WRe26
0.0 to 2400.0°C
0 to 400°C
±15.0°C
400 to 2400°C
±(0.2% of rdg + 2.0°C)
Type N (AWG14) 0.0 to 1300.0°C
±(0.2% of rdg + 1.3°C)
XK GOST
–200.0 to 600.0
–200 to –100°C
±(0.25% of rdg + 1.0°C)
–100 to 600°C
±(0.25% of rdg + 0.8°C)
Ni100 (SAMA)
–200.0 to 250.0°C ±(0.15% of rdg + 0.4°C)
Ni100 (DIN)
–60.0 to 180.0°C
±(0.15% of rdg + 0.4°C)
Ni120
–70.0 to 200.0°C
±(0.15% of rdg + 0.4°C)
RTD
1
Pt100 GOST
–200.0 to 600.0°C ±(0.15% of rdg + 0.3°C)
Cu100 GOST
–200.0 to 200.0°C ±(0.15% of rdg + 0.3°C)
Cu50 GOST
–200.0 to 200.0°C ±(0.4% of rdg + 0.5°C)
Pt200 (WEED)
–100.0 to 450.0°C ±(0.3% of rdg + 0.6°C)
1 Measuring current i = 1 mA
Input source resistance
Thermocouple input: 2 kΩ or less
RTD input:
1 Ω or less per wire (The resistance of all three wires must be equal).
Ambient temperature influence (with temperature variation of 10°C)
TC input
±(0.1% of rdg + 0.05% of range) or less, excluding the error of reference junction compensation
RTD input
±(0.2% of range + 2 digits) or less
Input source resistance
TC input
With variation of +1 kΩ: ±10 µV or less
RTD input
With variation of 1 Ω per wire (resistance of all three wires must be equal): ±(0.1% of rdg + 1 digit)
or less
With maximum difference of 100 mΩ between wires: Approx. 1 °C
Remote Control (/R1)
Item
Specifications
Number of input terminals
8
Input type
Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input
terminals, and shared common.
Input type and signal level
Voltage-free contact
Contact closed at 200 Ω or less and contact open at 100 kΩ or greater.
Open collector
ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 mA or less
Allowable input voltage
5 VDC
Signal type
Level or edge (250 ms or more)
Action
Executes a specified action by applying a given signal to the remote signal input terminal.
Action assignment: Set using the event action function
13.5 Options