Standard Performance
Item
Specifications
Measurement/display accuracy
Standard operating conditions:
Temperature:
23 ± 2°C
Humidity:
55 ± 10%RH
Power supply voltage:
90 to 132 or 180 to 250 VAC
Power supply frequency: 50/60 Hz ± 1%
Warm-up time:
At least 30 minutes.
Other ambient conditions such as vibration should not adversely affect the operation.
Input Type
Range
Measurement Accuracy (Digital display)
Digital Display
Max. Resolution
DC voltage
20 mV
±(0.05% of rdg + 12 digits)
1 μV
60 mV
±(0.05% of rdg + 3 digits)
10 μV
200 mV
10 μV
1 V
100 μV
2 V
±(0.05% of rdg + 12 digits)
100 μV
1 to 5 V
±(0.05% of rdg + 3 digits)
1 mV
6 V
1 mV
20 V
1 mV
50 V
10 mV
Thermocouple
(Not including
the accuracy of
reference junction
compensation; when
the burnout detection
function is off.)
R
±(0.15% of rdg + 1°C)
R, S
0 to 100°C: ±3.7°C,
100 to 300°C: ±1.5°C
B
400 to 600°C: ±2°C; accuracy not guaranteed for
temperatures less than 400°C
0.1°C
S
B
K
±(0.15% of rdg + 0.7°C)
–200 to –100°C: ±(0.15% of rdg + 1°C)
E
±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)
J
T
N
±(0.15% of rdg + 0.7°C)
–200 to 0°C: ±(0.35% of rdg + 0.7°C)
Accuracy not guaranteed for temperatures less than –200°C.
W
±(0.15% of rdg + 1°C)
L
±(0.15% of rdg + 0.5°C)
–200 to –100°C: ±(0.15% of rdg + 0.7°C)
U
WRe
±(0.2% of rdg + 2.5°C)
0 to 200°C: ±4.0°C
RTD
Pt100
±(0.15% of rdg + 0.3°C)
JPt100
DI
Voltage
Threshold level (Vth=2.4 V) accuracy ± 0.1 V
Contact
With parallel capacitance of 0.01 μF or less, 1 kΩ or less: 1 (ON). 100 kΩ or more: 0 (OFF)
Measuring accuracy in case of scaling
Accuracy during scaling (digits) = measurement accuracy (digits) × multiplier + 2 digits (rounded up)
* Fractions rounded up
where the multiplier = scaling span (digits)/measuring span (digits).
Example For 1-5 V range (A/D integration time is 16.7 ms or more), measurement span of 1.000 to
5.000 V, and scaling span of 0.000 to 2.000
The measuring accuracy for 5 V input is as follows.
Measuring accuracy (1-5 Vrange) = ±(0.05% × 5 V + 3 digits) = ±(0.0025 V [3 digits] + 3
digits) = ±6 digits
Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to 5.000) = 0.5
Thus, accuracy during scaling = ±(6 × 0.5 + 2) digits = 5 digits (rounded up)
Reference junction compensation accuracy
When measuring temperature greater than or equal to 0 °C and when input terminal temperature is
balanced
Type R, S, W, WRe: ±1.0°C
Type K, J, E, T, N, L, and U: ±0.5°C
Type B: Internal reference compensation is fixed to 0°C
13.6 General Specifications