Other communication errors – Yokogawa Value Series FX1000 User Manual
Page 238
10-13
IM 04L21B01-01EN
Troubleshooting
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
•
Maintenance and Test Communication Command Errors
An English error message is returned via the communication interface. These messages
are not displayed on the FX.
Code
Message
390
Command error.
391
Delimiter error.
392
Parameter error.
393
No permission.
394
No such connection.
395
Use ‘quit’ to close this connection.
396
Failed to disconnect.
397
No TCP control block.
398
Format error.
•
Other Communication Errors
An English error message is returned via the communication interface. These messages
are not displayed on the FX.
Code
Message
400
Input username.
Enter the name of a user that is registered on the FX and that is allowed to log into the FX through communication
commands.
401
Input password.
Enter the password.
402
Select username from ‘admin’ or ‘user’.
Select “admin” or “user” to specify the user level to connect to.
403
Login incorrect, try again!
The login failed. Enter the information again, starting with the user name.
404
No more login at the specified level is acceptable.
Connecting would cause the maximum number of simultaneous connections (administrator: 1, normal user: 2 to be
exceeded. Connect to a different level, or exit by entering “quit.”
420
Connection has been lost.
The connection has been lost.
421
The number of simultaneous connection has been exceeded.
The maximum number of simultaneous connections has been reached, so no further connections can be made.
422
Communication has timed-out.
Communication timed out so the connection was closed.
10.1 A List of Messages