beautypg.com

Digital display, Alarm indication – Yokogawa Value Series FX1000 User Manual

Page 23

background image

1-14

IM 04L21B01-01EN

• Alarm Indication

Alarm mark, alarm type, and measured value are displayed as follows according to

the alarm status.

Alarm

Release

Red

None

Red

None Red

Red

None None

None None

None

Alarm ACK

Alarm ACK

Occurrence

When indication is

set to non-hold

When indication is

set to hold

Alarm type

Red

Blue

Red

Blue Red

Red

Blue Blue

Blue Blue

Blue

Measured

value

Blinking

red

Blinking

green

Blinking

red

Green Red

Red

Green

Green

Green

Green

Green

Alarm mark

Digital Display

Displays the measured data numerically using large numbers. For the operating
procedure, see section 4.2.

Unit

Tag or channel number

Alarm mark

Measured value

Note

• Numeric display of measurement channels
If a measured value of a measurement channel is over range (see below), the measured

value is indicated as “+Over” or “–Over.” If a burnout is detected on a channel whose
burnout detection function is enabled, the word “Burnout” is indicated. Otherwise, a numeric
value is displayed.

Over range of measurement channels
• For DC voltage input, over range occurs when the measured value of the measurement

channel exceeds ±5% of the measurable range. For example, the measurable range
when the measurement range is 2 V is –2.000 to 2.000 V. If the measured value exceeds
2.200 V, + over range occurs; if the measured value falls below –2.200 V, – over range
occurs.

• For thermocouple or RTD input, over range occurs when the measured value exceeds

approximately ±10°C of the measurable range. For example, the measurable range
when the measurement range is R is 0.0 to 1760.0°C. If the measured value exceeds
approximately 1770.0°C, + over range occurs; if the measured value falls below
approximately –10.0°C, – over range occurs.

• For channels that are linearly scaled, + over range occurs when the value exceeds

30000 excluding the decimal point; – over range occurs when the value falls below
–30000. However, + over range can be changed to greater than or equal to 105% of the
scale width and – over range to less than or equal to –5% of the scale width within ±
30000.

For the setting procedure, see section 3.11.

• Numeric display of computation channels
See section 1.8, “Computation and Report Functions (/M1, /PM1, and /PWR1 options).”

1.3 Display