13.6 General Specifications

Construction

Item

Specifications

Mounting

Flush panel mounting (on a vertical plane)

Mounting angle

Inclined backward up to 30 degrees from a horizontal plane.

Allowable panel thickness

2 to 26 mm

Material

Case: Metal plate

Bezel and display cover:

Polycarbonate

Color

Case: Grayish blue green (Munsell 2.0B5.0/1.7 or equivalent)

Bezel: Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)

Front panel

Water and dust proof: Complies with IEC529-IP65, except side-by-side mounting

External dimensions

144 (W) × 144 (H) × 161.7 (D) mm (D: depth from the panel mounting plane)

Weight

FX1002, FX1004, FX1006: 1.3 kg

FX1008, FX1010, FX1012: 1.4 kg

(Not including options)

Normal Operating Conditions

Item

Specifications

Supply voltage

90 to 132, 180 to 250 VAC

Power supply frequency

50 Hz ± 2%, 60 Hz ± 2%

Ambient temperature

0 to 50°C

Ambient humidity

20 to 80%RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C)

Vibration

10 to 60 Hz, 0.2 m/s

2

Shock

Not allowed

Magnetic field

400 A/m or less (DC and 50/60 Hz)

Noise

Normal mode (50/60 Hz)

DC voltage

The peak value including the signal must be less than 1.2 times the measuring range.

Thermocouple

The peak value including the signal must be less than 1.2 times the measuring thermal electromotive force.

RTD

50 mV or less

Common mode noise

250 VACrms or less for all ranges (50/60 Hz)

Maximum noise voltage between channels

FX1xxx-x-x-H: 250 VACrms (50 or 60 Hz) or less

FX1xxx-x-x-L: 60 VACrms (50 or 60 Hz) or less

Mounting position

Can be inclined up to 30 degrees backward. Left and right horizontal.

Warm-up time

At least 30 minutes after power on

Installation location

Indoors

Operating altitude

2000 m or less

Power Supply

Item

Specifications

Rated supply voltage

100 to 240 VAC

Allowable power supply voltage range

90 to 132, 180 to 264 VAC

Rated power supply frequency 50 Hz, 60 Hz
Power consumption

Supply voltage LCD backlight off

Normal

Maximum

100 VAC 10 VA

15 VA

35 VA

240 VAC 15 VA

20 VA

45 VA

Allowable interruption time Less than 1 cycle of the power supply frequency