Sample mode, Real-time sample, Equivalent sample – RIGOL DS6000 Series User Manual

Page 87: Sample mode -5, Real-time sample -5 equivalent sample -5

background image

4 To Set the Sample System

RIGOL

DS6000 User’s Guide

4-5

Sample Mode


Press ACQSampling and press this softkey continuously to switch the sample
mode (the default is real-time sample). Note: all the Acquisition Mode support
real-time sample and equivalent sample.

Real-time Sample


In this mode, the oscilloscope samples and displays waveform within a trigger event.
The maximum real-time sample rate of DS6000 is 5 GSa/s and the current sample
rate is displayed under the Sa Rate menu.

Equivalent Sample


In this mode, the oscilloscope performs multiple trigger samples, samples at
different positions of each trigger and uses the sample points from multiple triggers
to rebuild the signal. The equivalent sample mode helps with detailed observation of
the repeated periodic signals. The maximum equivalent sample rate of DS6000 is
100 GSa/s.

Tip


Press Run/Stop to stop the sample, the oscilloscope will hold the last display. At
this point, you can still use the vertical control and horizontal control to pan and
zoom the waveform.