15 Troubleshooting


The commonly encountered failures and their solutions are listed below. When you
encounter those problems, please solve them following the corresponding steps. If
the problem remains still, please contact

RIGOL and provide your device

information (acquisition method: UTIL SystemSystem Info).

1. The screen is still dark (no display) after power on:

(1) Check whether the power is correctly connected or if the battery is correctly

installed.

(2) Check whether the power switch is really on.
(3) Restart the instrument after finishing the above inspections.
(4) If it still does not work correctly, please contact

RIGOL.

2. The signal is sampled but no waveform of the signal is displayed:

(1) Check whether the probe is correctly connected to the signal connecting

wire.

(2) Check whether the signal connecting wire is correctly connected to the BNC

(namely channel connector).

(3) Check whether the probe is correctly connected to the item under test.
(4) Check whether there are signals generated from the item under test (you

can connect the probe compensation signal to the problematic channel to
determine which has problems, the channel or the item under test).

(5) Resample the signal.

3. The tested voltage amplitude is greater or lower than the actual value

(note that this problem usually occurs when probe is used):
Check whether the attenuation coefficient of the channel complies with the
attenuation ratio of the probe.

4. There is waveform display but not stable:

(1) Check the trigger signal source: check whether the Source item at the

front panel complies with the signal channel actually used.

(2) Check the trigger type: general signals should use “Edge” trigger and video

signal should use “Video” trigger. Only when the proper trigger type is used,