beautypg.com

Horizontal reference, Horizontal reference -10 – RIGOL DS6000 Series User Manual

Page 80

background image

RIGOL

3 To Set the Horizontal System

DS6000 User’s Guide

3-10

Horizontal Reference


Horizontal reference is the reference position according to which the screen
waveform expands and compresses horizontally when adjusting HORIZONTAL
SCALE
. In Y-T mode, press MENUHorRef in the horizontal control area
(HORIZONTAL) to select the desired reference mode and the default is “Center”.

1. The Center of the Screen

When changing the horizontal time base, the waveform expands or compresses
horizontally around the center of the screen.


2. Trigger Position

When changing the horizontal time base, the waveform expands or compresses
horizontally around the trigger point.

3. User-defined

When changing the horizontal time base, the waveform expands or compresses
horizontally around the user-defined reference position (after selecting “User”,
use the knob to set the reference position and the range is from -350 to 350). In
the horizontal direction, the screen can display a maximum of 700 points with
the leftmost as 350 and the rightmost as -350. For example: set the reference
position to 150: