beautypg.com

Voltage parameters – RIGOL DS6000 Series User Manual

Page 142

background image

RIGOL

6 To Make Measurements

DS6000 User’s Guide

6-18

Voltage Parameters



1. Vmax: the voltage value from the highest point of the waveform to the GND.
2. Vmin: the voltage value from the lowest point of the waveform to the GND.
3. Vpp: the voltage value from the highest point to the lowest point of the

waveform.

4. Vtop: the voltage value from the flat top of the waveform to the GND.
5. Vbase: the voltage value from the flat base of the waveform to the GND.
6. Vamp: the voltage value from the top of the waveform to the base of the

waveform.

7. Vavg: the arithmetic average value on the whole waveform or on the gating

area.

n

x

Average

i

=

, where,

i

x

is the i

th

point being measured,

n

is the number of

points being measured.

8. Vrms: the root mean square value on the whole waveform or the gating area.

n

x

RMS

n

1

i

2

i

=

=

, where,

i

x

is the i

th

point being measured,

n

is the number of

points being measured.

Preshoot

Overshoot

Vamp

Vbase

Vmax

Vmin

Vpp

Vtop