beautypg.com

Video trigger, Video trigger -14 – RIGOL DS6000 Series User Manual

Page 106

background image

RIGOL

5 To Trigger the Oscilloscope

DS6000 User’s Guide

5-14

Video Trigger


Trigger on the standard video signal fields and lines of NTSC (National Television
Standards Committee), PAL (Phase Alternating Line), SECAM (sequential color with
memory) or HDTV (High Definition Television).

Trigger Type:
Press Type to select “Video”. At this point, the trigger setting information as shown
in the figure below is displayed at the upper right corner of the screen.


Source Selection:

Press Source to select CH1-CH4 as the Trigger Source. The current trigger source

is displayed at the upper right corner of the screen.
Note: select channel with signal input as trigger source to obtain stable trigger.

Video Polarity:
Press Polarity to select the desired video polarity. The polarities available are normal
polarity ( ) and inverted polarity ( ).

Sync:
Press Sync to select the desired sync type.
 All lines: trigger on all the horizontal sync pulses.
 Line number: for NTSC and PAL/SECAM video standards, trigger on the specified

line in the odd or even field; for HDTV video standard, trigger on the specified

line. Note that when this sync trigger mode is selected, you can modify the line

number using

in the Line Num menu with a step of 1. The range of the line

number is from 1 to 525 (NTSC), 1 to 625 (PAL/SECAM), 1 to 525 (480P), 1 to

625 (576P), 1 to 750 (720P), 1 to 1125 (1080P) or 1 to 1125 (1080I).

 Odd field: trigger on the rising edge of the first ramp waveform pulse in the odd

field.

 Even field: trigger on the rising edge of the first ramp waveform pulse in the

even field.

Video Standard:

Press Standard to select the desired video standard.
 NTSC: the field frequency is 60 fields per second and the frame frequency is 30

frames per second. The TV sweep line is 525 with the even field goes first and

the odd field follows behind.