To save the test mask, To save the test mask -4 – RIGOL DS6000 Series User Manual

Page 190

background image

RIGOL

9 Pass/Fail Test

DS6000 User’s Guide

9-4

Press Stop On Outp to select “ON” or “OFF”.
 ON: when failed waveforms are detected, the oscilloscope will stop the test and

enter the “STOP” state. At this point, the results of the test remain the same
with

the display on the screen (if display is turned on) and the [Trig

Out/Calibration] (if enabled) connector at the rear panel only outputs one
pulse.

 OFF: the oscilloscope will continue with the test even though failed waveforms

are detected. The test results on the screen will update continuously and the
[Trig Out/Calibration] connector at the rear panel will output a pulse each
time a failed waveform is detected.

Press Output to select “Fail” or “Fail+

”.

 Fail: when failed waveforms are detected, there are display and output but the

beeper does not sound.

 Fail+

: when failed waveforms are detected, there are display and output and

the beeper sounds (not related to the on/off state of the sound).

Press AuxOutput to quickly turn “ON” or “OFF” the measurement output from the
[Trig Out/Calibration] connector at the rear panel. You can also press UTIL
AuxOutput and select “PassFail” to set this output.

To Save the Test Mask


Users can save the current test mask to the Flash memory inside the instrument or
external USB storage device. The file format of the test mask file is “*.pf”. The
internal memory can store at most 10 test mask file (LocalPF.pf).

Press Save to enter file saving interface. Please refer to the relative description in
Store and Recall to save the test mask file to the internal or external memory.