10 Waveform Recording

DS6000 User’s Guide

10-8

again to restart the analysis.

5. Previous

After the waveform analysis finishes, press Previous can locate the error frame
before the current error frame. Pressing Run/Stop can also perform the
operation.

6. Next

After the waveform analysis finishes, press Next can locate the error frame
following the current error frame. Pressing Single can also perform the
operation.

7. Current Frame

Press Current Frame to adjust the data frame currently displayed and the
adjustable range is from 1 to the total number of frames in this waveform
analysis.
Note: you can use the navigation knob to view each frame of waveform of the
analysis.

8. Setup

Press Setup to open the detailed setting menu.
 Screen Start: press this key to set the start point of waveform analysis and

the range is from 5 to 685. The start point must be lower than the “Screen
End-10” currently set.

 Screen End: press this key to set the end point of waveform analysis and

the range is from 15 to 695. The end point must be greater than the
“Screen Start+10” currently set.

 Start Frame: press this key to set the start frame of waveform analysis and

the default is the first frame.

 End Frame: press this key to set the end frame of waveform analysis and

the default is the last frame.

 Threshold: press this key to set the threshold of waveform analysis and the

range is from 1% to 99%. The threshold is used to judge whether the frame
is an error frame. A frame can be recognize as an error one if the (relative)
difference between the frame and the templet is equal to or larger than the
threshold currently set.