2 Installation Tool


PH2 Phillips Screwdriver (recommended).

3 Installation Space


The following requirements must be fulfilled by the machine cabinet in which the instrument
is mounted.

The machine cabinet must be a standard 19-inch one.

At least 5U (222.2 mm) height should be provided by the machine cabinet.

The depth inside the machine cabinet should not be less than 400 mm.


The dimensions of the instrument after being installed are shown below (the unit is mm).


4 Installation Procedures


Only authorized operators can execute the installation operation. Improper installation might
result in damage of the instrument or incorrect installation of the instrument on the rack.

(1). Install the side plates: fix the side plates (left and right) onto the bottom plate of the

rack using four M4 screws.

(2). Install the front plate: fix the front plate onto the side plates (left and right) of the rack

using six M4 screws.

(3). Place the instrument: place the instrument by aligning it with the edges of the side

plates (left and right) and front plate of the rack and fix the instrument onto the bottom
plate through presser feet using four M4 screws.