16 Specifications

RIGOL

DS6000 User’s Guide

16-1

16 Specifications



All the specifications are guaranteed except the parameters marked with “Typical”
and the oscilloscope needs to operate for more than 30 minutes under the specified
operation temperature.

Sample

Sample Mode

Real-time Sample, Equivalent Sample

Real Time Sample Rate

5 GSa/s (single-channel)
2.5 Gsa/s (dual-channel)

Equivalent Sample Rate 100 Gsa/s

Peak Detect

200 ps (single-channel)
400 ps (dual-channel)

Averaging

After all the channels finish N samples at the same time,
N can be 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048,
4096 or 8192.

High Resolution

12 bits of resolution when ≥5 μs/div @ 5 GSa/s (or ≥10
μs/div @ 2.5 GSa/s).

Memory Depth

single-channel: Auto, 14k pts, 140k pts, 1.4M pts, 14M
pts and 140M pts are available
dual-channel: Auto, 7k pts, 70k pts, 700k pts, 7M pts and
70M pts are available

Input

Number of Channels

DS6XX4: four channels
DS6XX2: two channels

Input Coupling

DC, AC or GND

Input Impedance

(1 MΩ±1%) || (14 pF±3 pF)
or 50 Ω±1.5%

Probe Attenuation
Coefficient

0.01X to 1000X, in 1-2-5 step