9 Pass/Fail Test

RIGOL

DS6000 User’s Guide

9-3

To Select Source


Press Source to select the channel (channel enabled from CH1 to CH4) to be tested.
During the test, the oscilloscope will judge whether each frame of waveform in the
source complies with the current test mask and those waveforms pass through the
mask area (blue area) is considered as failed.

Mask Range


Users can define their desired test masks.

Press MaskRangeMaskRange to select “Screen Region” or “Cursor Region” for
test. When “Cursor Region” is selected, two gray cursor lines appear on the screen.
At this point, press Cursor A and Cursor B; then, use

to adjust the positions of

the two cursor lines respectively to determine the test range. Or, press CursorAB
and use

to adjust the positions of cursor A and curcor B at the same time. Note

that you can press down

continuously to switch the current cursor.


Press X Mask and Y Mask respectively and turn

, the mask lines appear on the

screen. Press Create Mask to apply the mask currently created. The ranges of the
horizontal and vertical adjustments are from 0.02 div to 4 div and 0.03 div to 4 div
respectively.

Test and Output


Before the test, you can use the following method to set the output mode of the test
results.
Press Msg Display to select “ON” or “OFF”. When “ON” is selected, the test results
will be displayed at the upper right corner of the screen.