6 To Make Measurements

RIGOL

DS6000 User’s Guide

6-31

Track Mode


In this mode, one or two pairs of cursors will appear. You can adjust the two pairs of
cursors to measure the X and Y values on two different sources respectively. The
points being measured on cursor A and B are marked by an orange rectangle and
rhombus respectively. When the cursors are moved horizontally, the marks will
position on the waveform automatically. When the waveform is expanded or
compressed horizontally, the marks will track the points being marked in the last
adjustment of the cursors.

Press CURSMode  “Track” to turn on the cursor track function and the
measurement results will be displayed at the upper left corner of the screen.


 The X value at cursor A (A->X): take the trigger position as reference and “s” or

“Hz” (when measuring FFT waveform) as its unit.

 The Y value at cursor A (A->Y): take the GND as reference and use the same

unit with the unit of the current source.

 The X value at cursor B (B->X): take the trigger position as reference and “s” or

“Hz” (when measuring FFT waveform) as its unit.

 The Y value at cursor B (B->Y): take the GND as reference and use the same

unit with the unit of the current source.

 The horizontal difference between cursor A and B (

△X).

 The reciprocal of the horizontal difference between cursor A and B (1/

△X).

 The vertical difference between cursor A and B (

△Y).


If needed, please refer to the following steps to modify track cursor measurement
parameters.