beautypg.com

Flexray decoding (option), Flexray decoding (option) -19 – RIGOL DS6000 Series User Manual

Page 177

background image

7 Protocol Decoding

RIGOL

DS6000 User’s Guide

7-19

FlexRay Decoding (Option)


Press MATHBUS1Decode and select “FlexRay” to open the FlexRay
decoding function menu.

1. Source

Press Source to select any channel (CH1 to CH4) as the signal source channel.

2. Signal Path

Press Signal Path to select the signal path (A or B) that matches the FlexRay
bus signal.

3. Signal Type

Press Signal Type to select the type of signal that matches the FlexRay bus.
The signal types available include BP, BM and TX/RX.

4. Baud

Press Baud to set the signal rate (2.5 Mb/s, 5Mb/s or 10 Mb/s) that matches the
FlexRay bus signal.

5. Sample Point

The sample point is the point within the bit time.The oscilloscope samples the bit
level at this point. The sample point position is expressed by the percentage of
the “time from the start of bit to the sample bit” takes up in “bit time”.
Press Sample Point and use

to adjust this parameter with a step of 1%

and the range is from 5% to 95%.


6. Threshold

Refer to the introduction in Parallel Decoding.

1 bit

60%

70%

80%