6 To Make Measurements

DS6000 User’s Guide

6-6

characteristics and are applicable to measure different waveforms. You need to
select the window function according to different waveforms and characteristics.
Press Window to select the desired window function and the default is
“Rectangle”.

Table 6-1 Window Function

Window

Characteristics

Waveforms Suitable for
Measurement

Rectangle The best frequency

resolution

The poorest amplitude

resolution
Similar to the situation
when no window is
multiplied

Transient or short pulse, the signal

levels before and after the

multiplication are basically the same;
Sine waveforms with the same
amplitude and rather similar
frequencies;
Wide band random noise with
relatively slowly changing waveform
spectrum.

Hanning

Better frequency

resolution

poorer amplitude

resolution

Sine, periodic and narrow band
random noise.

Hamming

A little bit better
frequency resolution
than Hanning

Transient or short pulse, the signal

levels before and after the

multiplication are rather different.

Blackman

The best amplitude
resolution
The poorest frequency
resolution

Single frequency signal, search for
higher order harmonics.


3. Set the Display Mode

Press Display to select “Split” (default) or “Full Screen” display mode.
Split: the source channel and the FFT operation results are displayed separately.
The time domain and frequency domain signals are straightforward.
Full Screen: the source channel and the FFT operation results are displayed in
the same window to view the frequency spectrum more clearly and to perform
more precise measurement.
Note: in FFT mode and when MATH is the active channel, you can also press
HORIZONTAL SCALE to switch to “Split” or “Full Screen”.