RIGOL DS6000 Series User Manual
Rp1010h高压示波器探头, Warning, Cleaning
用 户 手 册
User’s Guide
© 2013 RIGOL Technologies, Inc. All Rights Reserved.
RP1010H高压示波器探头
High Voltage Probe for Oscilloscope
北京普源精电科技有限公司
RIGOL Technologies,Inc.
邮箱
E-mail:[email protected] 网址Website:www.rigol.com
Warning
Do not attempt to take measurements from sources when
the chassis or return lead is not ground.
This ground connection is critical to the safety operation of
the probe. Failure to make this connection may result in
personal injuries or damages to the probe or voltmeter. This
connection must be made before the probe tip comes into
contact with the high voltage and must not be removed
until after the probe tip has been removed.
Do not connect the ground clip to the high voltage source or
the probe tip to the ground for any reason.
BEFORE turning the high voltage on, make sure that no
part of your body is in contact with the device.
Remembering that the voltage being measured is 1000
times greater than the voltmeter reading.
When the measuring voltage are DC 8 to 10 KV, or
AC rms 5 to 7 KV or AC peak 15 to 20 KV, do not
exceed 60 sec on each measurement. You can break off for
5 minutes at least during the measurement. Otherwise
overheating will affect the measurement accuracy even
damage the probe. Disconnect the probe tip from the high
voltage source BEFORE removing the ground clip lead.
Clean only the exterior probe body and cables. Use a soft
cotton cloth lightly moistened with a mild solution of
detergent and water. Do not submerge any portion of the
probe at any time.
Dry the probe thoroughly before any voltage measurement.
Make sure the probe is free from any solvents or solvent
fumes as these can case deterioration to the probe body and
cables.
When the measuring voltage is under
the followings. Each measurement
should be within 60 sec. You can break
off for 5 minutes at least during the
measurement.
DC: 8 ~ 10 KV
AC RMS: 5 ~ 7 KV
AC PEAK: 15~ 20 KV
当测量电压为下述情况时,每
次测量时间不得超过60秒。期
间需断开并等待散热至少5分
钟。
直流:
8 ~ 10 KV
交流
有效值:5 ~ 7 KV
交流
峰峰值:15 ~ 20 KV
!
Cleaning
性能指标
(Specifications)
最大工
作电压
Max. Working Voltage
DC: 0 ~ 10 KV DC
AC: pulse ≤ 20 KV peak to peak
AC: sine wave ≤ 7 KV rms
信噪比
Signal/Noise
DC ≥ 60 dB (1 KHz),
≥
50 dB (1 MHz)
带宽
Bandwidth
DC ~40 MHz (-3 dB)
衰减比例
Division Ratio
1:1000
上升时间
Rise Time
≤
8.8 ns
输入阻抗
Input Impedance
100 MΩ ± 5%
输入电容
Input Capacitance
3.0 pF ± 0.5 pF
补偿范围
Compensation Range
5 pF ~ 50 pF
线长度
Cable Length
2.0 meter (± 0.2 M)
温度系数
Temperature Coefficient ≤ 200 PPM /℃
精确度
Accuracy
DC
±
3% (DC to
10 KV)
AC
±
3% (1 KHz/1 KV )
-3dB 40 MHz
安全认证
Safety App.
CE
操作温度
Operating Temperature 0 ~ 50℃
存储温度
Storage Temperature
-20 ~ +70℃
重量
/尺寸 Weight / Dimension
250 g /
Φ
75 x 340
L
mm
S M1 0 2 0 0 0 2 5 7 9