5 To Trigger the Oscilloscope

RIGOL

DS6000 User’s Guide

5-27

Sample Point:
Sample point is the point within the bit time. The oscilloscope samples the bit level at
this point. The sample point position is represented by the percentage that the “time
from the start of the bit to the sample point” takes up in the “bit time”.
Press SamplePoint and use

to modify the parameter with the step as 1% and

the range is from 5% to 95%.








Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,

normal or single. The corresponding status light of the current trigger mode turns on.

Trigger Setting:
Press Setting to set the trigger parameter (trigger holdoff) under this trigger type.

Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of “Trigger level” on page 5-8.


1Bit

60%

70%

80%