Measurement Setting
1. Source Selection
Press MEAS Source to select the desired channel for measurement
(CH1-CH4 or MATH). The color of the parameter icon under MENU at the left of
the screen will change with the source selected.
2. Measurement Range
Press MEAS Meas.Range Meas.Range to select “Screen Region” or
“Cursor Region” for measurement. When “Cursor Region” is selected, two cursor
lines appear on the screen. At this point, press Cursor A and Cursor B and use
to adjust the positions of the two cursor lines respectively so as to determine
the measurement range. Or, press CursorAB and use
to adjust the
positions of cursor A and cursor B at the same time. Note: you can press
continuously to switch among the cursors.
3. Delayed Measurement Setting
Specify the source A and source B in the measurement items “Delay A→B ”
and “Delay A→B ”.
Press MEAS Meas Setting Type “Delay” and then press Source A
and Source B to set the two channel sources (CH1-CH4) of the delayed
measurement respectively.