6 To Make Measurements

RIGOL

DS6000 User’s Guide

6-21

Measurement Setting


1. Source Selection

Press MEASSource to select the desired channel for measurement
(CH1-CH4 or MATH). The color of the parameter icon under MENU at the left of
the screen will change with the source selected.

2. Measurement Range

Press MEASMeas.RangeMeas.Range to select “Screen Region” or
“Cursor Region” for measurement. When “Cursor Region” is selected, two cursor
lines appear on the screen. At this point, press Cursor A and Cursor B and use

to adjust the positions of the two cursor lines respectively so as to determine

the measurement range. Or, press CursorAB and use

to adjust the

positions of cursor A and cursor B at the same time. Note: you can press

continuously to switch among the cursors.


3. Delayed Measurement Setting

Specify the source A and source B in the measurement items “Delay AB
and “Delay AB ”.
Press MEASMeas SettingType  “Delay” and then press Source A
and Source B to set the two channel sources (CH1-CH4) of the delayed
measurement respectively.