Horizontal, Vertical – RIGOL DS6000 Series User Manual

Page 268

background image

RIGOL

16 Specifications

DS6000 User’s Guide

16-2

Maximum Input

Voltage (1MΩ)

Maximum Input Voltage of the Analog Channel
CAT I 300 Vrms, CAT II 100 Vrms,
Transient Overvoltage 1000V pk
with RP2200 10:1 probe: CAT II 300 Vrms
with RP3300 10:1 probe: CAT II 300 Vrms
with RP3500 10:1 probe: CAT II 300 Vrms
with RP5600 10:1 probe: CAT II 300 Vrms



Horizontal

Timebase Scale

DS606X: 1 ns to 1000 s
DS610X: 500 ps to 1000 s

Timebase Accuracy

≤ ±(5 + 2 ×instrument age in years) ppm

Delay Range

Pre-trigger (negative delay): ≥1 screen width
Post-trigger (positive delay): 1 s to 100,000 s

Timebase Mode

Y-T, X-Y, Roll, Delayed Sweep

Number of XYs

2 simultaneously

Waveform Capture
Rate

[1]

150,000 wfms (vector display); 180,000 wfms (dots
display)

Vertical

Bandwidth (-3dB)

DS606X: DC to 600 MHz
DS610X: DC to 1 GHz

Single-shot Bandwidth DS606X: DC to 600 MHz

DS610X: DC to 1 GHz (each channel)

Vertical Resolution

8bit, two channels sample at the same time

Vertical Scale

2 mV/div to 5 V/div (1 MΩ)
2 mV/div to 1 V/div (50 Ω)

Offset Range

2 mV/div to 124 mV/div: ± 1.2 V (50 Ω)
126 mV/div to 1 V/div: ± 12 V (50 Ω)
2 mV/div to 225 mV/div: ± 2 V (1 MΩ)
230 mV/div to 5 V/div: ± 40 V (1 MΩ)

Bandwidth Limit

[2]

20 MHz or 250 MHz