RIGOL DS6000 Series User Manual
Page 153
RIGOL
DS6000 User’s Guide
6-29
3. Select X (Y) Axis Unit
When the cursor type is “Time”, press Time Unit to select “s”, “Hz”, “°” or “%”.
S: when this unit is selected, in the measurement results, the CurA, CurB
and △X are in “s” and 1/△X is in “Hz”.
Hz: when this unit is selected, in the measurement results, CurA, CurB and
△X are in “Hz” and 1/△X is in “s”.
°: when this unit is selected, in the measurement results, CurA, CurB and
△X are in “°”. At this point, CurA, CurB and △X will change to “0°”, “360°”
and “360°” respectively when you press Set Cursor, no matter where
cursor A and B are currently located. At the same time, two cursor lines
(unmovable) appear on the screen as the reference positions.
%: when this unit is selected, in the measurement results, CurA, CurB and
△X are in “%”. At this point, CurA, CurB and △X will change to “0%”,
“100%” and “100%” respectively when you press Set Cursor, no matter
where cursor A and B are currently located. At the same time, two cursor
lines (unmovable) appear on the screen as the reference positions.
When the cursor type is “Voltage”, press Vertical Unit to select “Source Unit” or
“%”.
Source Unit: when this unit is selected, in the measurement, the units of
CurA, CurB and △Y will automatically be set to the unit of the current
source.
%: when this unit is selected, in the measurement results, CurA, CurB and
△Y are in “%”. At this point, CurA, CurB and △Y will change to “0%”,
“100%” and “100%” when you press Set Cursor, no matter where cursor
A and B are currently located. At the same time, two cursor lines
(unmovable) appear on the screen as the reference positions.
4. Adjust the Cursors Position (note that under the same menu page,
you can also press
continuously to switch the current cursor)
Adjust cursor A: press CursorA and use
to adjust the position of cursor
A. During the adjustment, the measurement result will change accordingly.
The adjustable range is limited within the screen.
Adjust cursor B: press CursorB and use
to adjust the position of cursor
B. During the adjustment, the measurement result will change accordingly.
The adjustable range is limited within the screen.
Adjust cursor A and B at the same time: press CursorAB and use
to