5 To Trigger the Oscilloscope

RIGOL

DS6000 User’s Guide

5-19

When this trigger condition is selected:
--press Data Bits to select “5 bit”, “6 bit”, “7 bit” or “8 bit”;
--press Data and input the data value according to the setting in Data Bits and
the upper limit can be 31, 63, 127 or 255.


Baud Rate:
Press Baud to select the baud rate (equal to specifying a clock frequency) of the
data transmission. The rates available include 2400 bps, 4800 bps, 9600 bps
(default), 19200 bps, 38400 bps, 57600 bps, 115200 bps and user-defined. After
“User” is selected, press Setup and use

to set more specific value from 1 to

900000 with the adjustment step as 1 bps.

Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,

normal or single. The corresponding status light of the current trigger mode turns on.

Trigger Setting:
Press Setting to set the trigger parameter (trigger holdoff) under this trigger type.

Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of “Trigger Level” on page 5-8.