5 To Trigger the Oscilloscope

DS6000 User’s Guide

5-10

Pulse Width Setting:
In this oscilloscope, positive pulse width is defined as the time difference between
the two crossing points of trigger level and positive pulse. As shown in the figure
below:






 When the Pulse Condition is set to

,

,

or

, press

Setting and use

to input the desired value. The range available is from 4 ns

to 4 s.

 When the Pulse Condition is set to

or

, press Upper Limit and

Lower Limit respectively and use

to input the desired values. The range of

the upper limit available is from 12 ns to 4 s. The range of the lower limit
available is from 4 ns to 3.99 s. Note: the lower limit of the pulse width must be
lower than the upper limit.


Trigger Mode:

Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,

normal or single. The corresponding status light of the current trigger mode turns on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger holdoff) under this trigger type.

Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of “Trigger Level” on page 5-8.



Trigger level

Positive

pulse width

A

B