5-10
Pulse Width Setting:
In this oscilloscope, positive pulse width is defined as the time difference between
the two crossing points of trigger level and positive pulse. As shown in the figure
below:
When the Pulse Condition is set to
,
,
or
, press
Setting and use
to input the desired value. The range available is from 4 ns
to 4 s.
When the Pulse Condition is set to
or
, press Upper Limit and
Lower Limit respectively and use
to input the desired values. The range of
the upper limit available is from 12 ns to 4 s. The range of the lower limit
available is from 4 ns to 3.99 s. Note: the lower limit of the pulse width must be
lower than the upper limit.
Trigger Mode:
Press Sweep to select the Trigger Mode (page 5-3) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff) under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of “Trigger Level” on page 5-8.
Trigger level
Positive
pulse width
A
B