6 To Make Measurements

DS6000 User’s Guide

6-28

Manual Mode


In this mode, a pair of cursors will appear. You can adjust the cursors to measure the
X (or Y), X increment (or Y increment), and reciprocal of the X increment on the
selected source (CH1-CH4 or MATH).

Press CURSMode  “Manual” to turn the manual cursor function on and the
measurement results will be displayed at the upper left corner of the screen in the
following mode.


 The X or Y value at cursor A (CurA): X value takes the trigger position as

reference and Y value takes the channel GND as reference.

 The X or Y value at cursor B (CurB): X value takes the trigger position as

reference and Y value takes the channel GND as reference.

 The horizontal or vertical difference between cursor A and B (

△X or △Y).

 The reciprocal of the horizontal difference between cursor A and B (1/

△X).


If needed, please refer to the following steps to modify the parameters of manual
cursor measurement.

1. Select Cursor Type

Press Type to select “Time” or “Voltage”. The time cursor is a pair of vertical
dotted lines and is usually used to measure time parameters. The voltage cursor
is a pair of horizontal dotted lines and is usually used to measure voltage
parameters.

2. Select Measurement Source

Press Source to select the waveform in the analog channels (CH1-CH4) or math
operation results (MATH) for measurement (only channels enabled can be
selected). If “None” is selected, no cursor will be displayed.