8.2.19 Still Water Zero Adjustment

Zero point adjustment must be performed in the state the fluid in the detector's measurement tube is

held still.

Key operation

Display example

Description

E:

Select "ZERO ADJUST" in the configuration item

MEAS MODE

^ZERO ADJUST

[SEL] [ENT]

selection screen.

STEP1

El:

The current flow rate measurement value is

ZERO ADJUST

01.0

displayed.

[ENT]

%

Then press the [SET] switch.

[EXT] [SET]

* Pressing the [EXT] key returns you to the menu

screen.

STEP2

El:

When you press the [SET] key, the title display

ADJUST READY

changes to ADJUST READY, enabling you to

01.0

perform adjustment.

[SET]

%

* Pressing the [EXT] key returns you to the previous

[EXT] [SET]

screen.

STEPS

El:

Hold down the [SET] key when ADJUST READY is

ZERO ADJUST

displayed causes "ZERO ADJUST" to be displayed

[SETjHold down

‘ZERO ADJUST*

and still water zero adjustment to start.

Switch operation is disabled during adjustment.

STEP4

El:

After approximately 30 seconds, still water zero

ZERO ADJUST

adjustment is completed and the flow rate

00.0

%

[EXT] [SET]

measurement value after adjustment is displayed.

Press the [EXT] to return to the menu screen.

Note 1; To start still water zero adjustment, hold down the [SET] key.
Note 2: Still water zero point adjustment is possible only when the flow rate value is within a

range of ±1.25 m/s.

Note 3: If you want to cancel the adjustment when ADJUST READY is displayed, press the

[EXT] key. This returns you to the flow rate measurement value display screen.

-