10.8 Converter Failure Alarm
When one or more of the following converter errors occur in a self-diagnostics sequence, an alarm
signal can be output. See Chapter 12, “Self-Diagnostics and Warning Functions” for details of each
alarm status.
■ Self-diagnostics errors
Self-diagnosi
(LCD dis
tics error
Play)_______________
Error contents
*
R
0
M
E
R
R
0
R
*
ROM error
*
R
A
M
E
R
R
0
R
♦
RAM error
P
A
R
A
M
E
T
E
R
System parameter error
F
A
1
L
U
R
E
E
X
C
U
R
R
E
N
T
Excitation circuit open or disconnection
0
P
E
N
E
X
C
U
R
R
E
N
T
Excitation current error, excitation circuit fault
E
R
R
0
R
A
D
C
E
R
R
0
R
ADC error
1
N
V
A
L
1
D
Invalid totalizer counts
T
0
T
A
L
* Error message is displayed in measurement sub screen.
If you want to use a converter error alarm output, set digital output 1 or 2 (DOl or D02) to the
error alarm output of the converter following 0, “Digital Input/Output.”
In addition, set the alarm output condition to normally open (NORMAL OPEN) or normally
close (NORMAL CLOSE) status.
Output conditions
• Normal Open; transistor / relay contact is closed when an error occurs.
• Normal Close; transistor/relay contact is open when an error occurs.
Note: Alarm output contacts are open while the converter is powered off.
-
132
-