O Mode switching
The electromagnetic flowmeter usually operates in the measurement mode.
When you need to perform parameter setting, calibration and adjustment, you enter the setting mode.
To enter the setting mode, hold down center key for 3 seconds or longer.
Holding down the desired key for 3 seconds or longer will display screen of code input. When "111"
was input with this screen, [SET] and [CNT] (for switch operation) are displayed in the measurement
screen. When the code except "111" was input, label is not displayed.
[SET] key
You enter the setting mode (the menu configuration setting screen is displayed).
[CNT] key
The screen is switched to the total count control screen, enabling you to
manipulate the total counter.
Note: If password has been set, the password input screen appears when you move from the
measurement mode to the menu configuration setting screen (when you press the [SET] key) and
when you move to the total count control screen (when you press the [CNT] key).
If the password you enter does not match, you will be unable to change some of the parameters.
Also you will be unable to clear the total count when you are manipulating the total counter.
O Function of operation time-out
When a converter is state of the setting mode, and there is not operation more than 1 minute, mode
returns to the measurement mode automatically.
Menu display
A:
MEAS MODE
-^EX CURRENT
METER SIZE
EX FREQ .
FLOW DIRECTN
PASSWORD
ADDR SET
[SELI [ENT]
Parameter confirmation
display
A1:
EX CURRENT
0.19000
A
[EXT] [ENT]
Parameter change display
A1:
EX CURRENT
0.1900
A
[UP] [SEL] [SET]
Mode returns to the
measurement mode
automatically
Mode does not
return to the
measurement mode
-