Electromagnetic Flowmeter LF600 Series
Operation Guide
ISetting Parameters Flow!
Items
Screen Display
Setting or selecting value
0 Measurement Mode
MEAS MODE
1 Exicitation Current
EX CURRENT
O.OOOOA ~ 0.2400A (0.0001A each)
2 Meter size
METER SIZE
0.1 24 inch. 2.5mm 600mm
A
3 Exicitation Freouencv
EX FREQ
6.12.24 Hz
4 Flow Direction
FLOWDIRCTN
NORMAL. SWITCH
5 Password
PASSWORD
000 ~ 999 (leach)
6 Communication Address
ADDR SET{*2)
000 126 (1 each)
0 Measurement Mode
MEAS MODE
1 Main Display
MAIN DSP
■First Unit : Volume, Others
%, m3,1, ml, bbl, gal, m/s, ft/s, COUNT, RANGE, pt, qt,
CUSTOM, GRAPH И)
■Second Unit ;Time unit
В
2 Sub Display
SUB DSP
■Third unit :Flow direction code
В (Automatic selection bi-directionflow)
F (Fixed forward flow), R (Fixed reverse flow)
D (Différence)
3 Custom Coefficient
CUSTOM DATA
0 99999999
4 Custom Unit
CUSTOM UNIT
7 Characters
5 LCD Density Adiustment
LCD ADJUST
1(LlEht)~5(DariO
6 Swich Position
SW POSITION
TOP, ВОПОМ. LEFT, RIGHT
0 Measurement Mode
MEAS MODE
1 Ranse Tvpe
RANGE TYPE
SINGLE. 4F - OR. 2F - 2R. EXT 2F - OR. EXT 2F - 2R
2 Range 1
RANGE1
■Volume unit
m3,1, ml, gal, bbl, pt, qt
■Time unit
C 3 Range 2
RANGE2
4 Range 3
RANGES
/s, /min, /h, /d
■Velocity unit
5 Range 4
RANGE4
m/s, ft/s
6 Ranee Hysteresis
RANGE HYS
0.0 ~ 25.0)1(0.1)1 each)
0 Measurement Mode
MEAS MODE
1 Dampine Constant
DAMPING
00.0.00.5.01.0 ~ 60.0 6
D
2 Low Cutoff
LOW CUT
0.0 ~ 10.0 «(О.« each)
3 4-20mA Alam Cutout
ALM mA SET
UNDER 3.0mA, 4.0mA. HOLD, OVER 24.0mA
4 Display Low Cutoff
DSP LOW CUT
ON. OFF
5 Output Low Limit
LOW LIMIT
4.0mA, 3.2mA, 2.4mA
E
0 Measurement Mode
MEAS MODE
1 Zero Adiustment
ZERO ADJUST
Hold down the [SET] kev
0 Measurement Mode
MEAS MODE
1 D01 Function
D01 FUNCTN
NO USE, HIGH ALM, HH ALM, LOW ALM, LL ALM
2 D02 Function
D02FUNCTN(*2)
EMPTY ALM. PULSE OUT. PULSE FRD. PULSE REV
F
3 Dl Function
Dl FUNCTN (*2)
NO USE. ONT STA/STP, CNT RES/STA
RANGE SW. ZERO ADJ. FIXED OUT
4 D01 Active Status
D01 ALM STS
5 D02 Active Status
D02ALMSTS(»2)
6 Dl Detective Level
Dl DETLV{*2)
H LEVEL, L LEVEL
0 Measurement Mode
MEAS MODE
G
1 Counting Rate
COUNT RATE
Units : m3,1, ml, bbl, gal, pt, qt
Value : within the range 3.6 ~ 36000000 pulse/h
(0.001 - 10,000 pps)
2 Pulse Mode
PLS MODE
AUTO. MANUAL
3 Pulse Width
PLS WIDTH
0.3 500.0ms (0.1ms each)
or Less than half of the pulse rate for 100% flow rate output
0 Measurement Mode
MEAS MODE
H 1 Preset Count
PRESET ONT
0 99999999 count (1 count each)
2 Preset Function
PRESET FNO
HOLD. 50ms PULSE. 500ms PULSE
0 Measurement Mode
MEAS MODE
1 High Alarm Set
HALM SET
ON. OFF
1 2 High Alarm Value
H ALM VAL
-10 ~ 110)1
3 Low Alarm Set
L ALM SET
ON. OFF
4 Low Alarm Value
L ALM VAL
-10 ~ 110)1
0 Measurement Mode
MEAS MODE
1 High High Alarm Set
HH ALM SET
ON. OFF
J
2 High High Alarm Value
HH ALM VAL
-10-110%
3 Low Low Alarm Set
LL ALM SET
ON. OFF
4 Low Low Alarm Value
LL ALM VAL
-10-110%
0 Measurement Mode
MEAS MODE
К
1 Empty Pipe Alarm
EMPTY ALM
OFF. NORMAL. SENSITIVE. SENSITIVE-H
2 SelNIagnosIs function
SELF CHECK
ON. OFF
3 Alarm Output Preset
ALM PRESET
WITHOUT EMP. WITH EMP
0 Measurement Mode
MEAS MODE
L
1 Rate-of-change Limit
LIMIT RATE
0.0 "30.0)1(0.1)1 each)
2 Control Limit Time
LIMIT TIME
00 — 20 s (1 s each)
0 Measurement Mode
MEAS MODE
м
1 Fixed-value Output
FIXED OUT
ON. OFF
2 Fixed-current Output
FIOED OURR
2.4" 24.0 mA (0.1mA each)
3 Fixed-pulse Output
FIXED PULSE
0 — 10000 pps (Ipps each)
N
0 Measurement Mode
MEAS MODE
1 Manual Zero
MANUAL ZERO
±10% of lOm/s-maximum range (0.1% each)
0 Measurement Mode
MEAS MODE
1 0% Flow Rate Calibration
FLOW CALO
[ENTl switch (Keep pushing more than 2 seconds^
0
2 50% Flow Rate Calibration
FLOW CAL 50
None (Displav)
3 100% Flow Rate Calibration
FLOW CAL 100
fENTl switch (Keep pushing more than 2 seconds.)
4 Checking the Excitation Current Value
EXCURRDSP
None (Display)
♦ 10
♦11
♦12
♦13
___ : Entering value
___ : Selecting value
I
: Entering and selecting value
*4
*5
*7
*9
*10
*11
*12
*13
4-20mA Alarm Output
Output Status
0: UNDER 3.0mA
1:4.0mA
2: HOLD
3: OVER 24.0mA
Under 3mA output
4mA output
Measured data hold
Over 24mA output
D01.D02 Items
Digital output function
0:NO USE
Not used
1: HIGH ALM
High limit alarm output
2:HHALM
High high limit alarm output
3:L0WALM
Low limit alarm output
4:LLALM
Low low limit alarm output
5:RNG SIG1
Multi - range output No.1
6:RNG SIG2
Multi - range output No.2
7:PRESET
Preset point output
8:CONVALM
Converter failure alarm output
9:EMPTYALM
Empty pipe alarm output
10:PULSE OUT
Pulse output (bldrectlonal flow)
11:PULSEFRD
Pulse output (fixed forward flow)
12:PULSE REV
Pulse output (fixed reverse flow)
This value is factory adjusted when shipped.
These functions are option.
case of choosing COUNT or RANGE
COUN : displays totalized flow counts (8 digits) without a unit
RANGE: displays the range number (1 to 4)_ _ _ _ _ _ _ _ _ _ _ _ _
Only sub displey
1: SINGLE
2:4F-OR
3:2F-2R
4: EXT. 2F-
5: EXT. 2F-
Description
Single range
Unidirectional flw, automatic selection of multiple ranges
Bidirectional flows, automatic selestion of multiple ranges
Unidirectional flows, multiple ranges selected by exterminai single
Bidirectional flows, multiple ranges selected by exterminai single
D01/D02 Active Status
Alarm - output Status
0:NORMAL CLOSE
1: NORMAL OPEN
Normal close
Normal open
Dl Detective Level
Detective Level
0:L LEVEL
1:H LEVEL
L level
H level
Preset Function
Output Status
0:HOLD
1:50ms PULSE
2:500ms PULSE
Output Status Level Hold
Pulse out (pulse width 50ms)
Pulse out (pulse width 500ms)
Empty Pipe Alarm
Setting and Degective Level
0:OFF
1: NORMAL
2:SENSITIVE
2:SENSITIVE-H
Not used
Used and Ditective level low
Used and Ditective level middle
Used and Ditective level high
When this function is set to ON, the conditions are foilwing.
Items
Conditions
Current output
Pulse output
Digital outputs
Data Indicating
User - set current output
Pulse output with a user - set counting rate
Previous status Is retained (excluding pulse output).
Instaneous flow rates and flow velocity (no totalization).
(1) In case of the zero adjustment (Menu No.EI),Zero offset is automatically cleared to Zero.
(2) Calculate the zero offset value with the foilwing equation:
Zero offset value (U) = ((Actual flow rate) - (LF600 measured value))
Measured condition
Flow rate
91 In measured span
Actual flow rate obtained
from other instrument
10.0 m3/min
5091
LF600 measured value
10.5 m3/min
52.591
Zero offset
-
-2.591
If zero offset is set to -2.5X, LF600 will output 50.091 flow rate instead of 52.591
LF600seriesGuide070207cs.indd 1-2