6|f|8ia0i8,6i9i, Toshiba – Toshiba LF600F User Manual
Page 100
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".
TOSHIBA
6|F|8iA0i8,6i9i
Changing the high, low, high-high, or low-low limit alarm value
Shown below is an example of changing the high limit alarm value from +105% to +103%.
Key operation
Display example
Description
I:
MEAS MODE
HALM SET
ALM VAL
LALM SET
L ALM VAL
[SEL] [ENT]
Select "HALM VAL" in the configuration item
selection screen.
STEP1
[ENT]
I2:
HALM VAL
+105.0
%
[EXT] [ENT]
The currently set value (+105.0 in this example) is
displayed.
Then press the [ENT] key.
* Pressing the [EXT] key returns you to the menu
screen.
STEP2
[ENT]
I2:
HALM VAL
B105.0
%
[U P] [SEL] [SET]
The switch name display at the bottom changes
(three keys [UP], [SEL] and [SET]).
At the same time, the cursor appears.
(The digit on which the cursor is positioned is
reverse-displayed.)
STEPS
[SEL]0[UP]
12:
H ALM VAL
+10|.0
%
[U P] [SEL] [SET]
Move the cursor to the desired digit using the [SEL]
key and change the numeric value using the [UP]
switch.
Repeat this operation to change the value to
+103.0%.
When you have selected the desired item, press
the [SET] key to set the value temporarily.
STEP4
[SET]
12:
H ALM VAL
+103.0
%
OK?
^
[NOl [SETl
When you press the [SET] key, a message
confirming you whether the setting is OK is
displayed.
If OK, press the [SET] key. If you want to redo the
setting, press the [NO] key.
STEP5-1
[NO]
12:
H ALM VAL
B105.0
%
[U P] [SEL] [SETl
Pressing the [NO] key when you are asked "OK?"
causes the numeric value to return to the previous
value and enables you to redo the setting.
STEP5-2(=END)
[SET]
12:
HALM VAL
+103.0
%
[EXT] [ENT]
Pressing the [SET] key when you are asked "OK?"
causes the setting to be determined.
Press the [EXT] key. You return to the menu
screen.
Pressing the [ENT] key enables you to restart
setting value change operation from a state of
+103.0%.
Note: If you set a value that is not within a range of -10% to +110%, * LOW OVER SPEC* or
♦HIGH OVER SPEC* is displayed and the value returns to the one before change. Set a
value once again.
-
99
-