10.7 Remote Selection of Fixed Value Output
A user-specifled 4-20 mA output and pulse output can be selected with a DI signal.
Proceed as follows to use this function:
Fixed-value setting
Set the fixed-value for current output and for pulse output. See 0, “Fixed-Value
Output.” Set the fixed-value output enable/disable status to “OFF.”
If the pulse output is not used, fixed-value setting for pulse output is not
needed.
DI function setting
Set DI to use as a fixed-value output control signal. See 0, “Digital I/O.”
Control signal input conditions:
Control signal input level
4 -20 mA and pulse output
L level
Outputs the measured value.
H level
Outputs the fixed-value.
-