9.2.2 50 % Flow Rate Calibration
Using the converter's internal calibration circuit, 50% flow rate calibration can be performed. For the
calibration procedure, refer to the calibration procedure for 0% flow rate calibration (Or select the
"FLOW CAL 50" from the menu).
9.2.3100 % Flow Rate (Span) Calibration
Using the converter's internal calibration circuit, 100% flow rate calibration can be performed. For the
calibration procedure, refer to the calibration procedure for 0% (Or select the "FLOW CAL 100" from
the menu.).
9.2.4 Checking the Excitation Current Value
You can monitor the exciting current value.
■ To check the exciting current value:
Key operation
Display example
Description
0:
MEAS MODE
FLOW CAL 0
FLOW CAL 50
FLOW CALI 00
^EXCURR DSP
iSELl [ENTl
Select "EX CURR DSP" in the configuration item
selection screen.
[ENT]
04;
EX CURR DSP
0.2000
A
The excitation current value is displayed.
Pressing the [EXT] key returns you to the menu
screen.
_________ lEXJ]_________
The excitation current value is factory adjusted when shipped. Contact you're nearest Toshiba
representative if any change is necessary.
-