5.4.2 Terminal Treatment of Cables

Follow the procedures below to treat the terminals (at the converter side) of various cables and install

the cables to the terminal block. Use appropriate cables based on the description in Section 5.1

"Cables." Crimp a round type insulated crimp-type terminal to the end of the cables.

(1) Power cable, current output cable, digital I/O cables

The necessary cables should be ordered from the person responsible for the installation. Strip the
sheath of each conductor as shown in Figure 5.6 and attach a crimping terminal with insulated
sleeve to it. The size of the crimping terminal is as follows:

Integral type LF600F: M4

Separate type LF602F: M3.5

• Connect the power cable to terminal blocks LI and L2.

• Connect the current output cable to terminal blocks + and -.

• Connect the digital I/O cable to terminal blocks Dl, DOl, D02 and COM, as required.

Xrimping terminal-f

^ ^

1LF602F type: M3.5

Figure 5.6 Terminal Treatment of Power Cable, Current Output Cable

and Digital I/O cable

LF602F

(2) Excitation cable

Strip the sheath from the end of each conductor as shown in Figure 5.7, attach an M3.5 crimping

terminal with insulated sleeve, and connect it to the terminal blocks X and Y. Connect the red
conductor to terminal block E.

Figure 5.7 Terminal Treatment of Excitation Cable

-