Preset count output performance
(l)The following is an example for totalizer flow counts output in which the totalizer is reset with an
external signal (when preset output status level hold mode is set (contact ON)).
H level input
Counter Reset/Start signal
(DI detective level is H).
Totalizer flow counts
n
Counter reset
Counter start
Counter reset
Preset point output
L level input
Preset point
ON
OFF
Input/Output signal time chart
* When the Reset/Start signal is in H level (DI counter control signal level: H), the totalizer
is reset to zero and stops counting. When the Reset/Start signal goes to L level, the
totalizer starts counting. The preset point output goes ON when the totalizer counts
reaches the preset point, and the output goes OFF when the totalizer is reset to zero.
(2)The following is an example for totalizer flow counts output in which the totalizer is reset with an
external signal (when one-shot pulse output mode is set).
H level input
L level input
Preset point
ON
OFF
the preset output goes OFF.
the output goes OFF
Input/Output signal time chart
* When the Reset/Start signal is in L level (DI counter control signal level: L), the totalizer is
reset to zero and stops counting. When the Reset/Start signal goes to H level, the totalizer
starts counting.
The preset point output goes ON when the totalizer counts reaches the preset point. The
output goes OFF when the totalizer is reset to zero or when it takes the time set pulse
width from the output goes ON.
-