4.0
SYSTEM PLUMBING
This chapter discusses recommended plumbing practices and provides general guidelines for system setup.
The system will work more efficiently if tubing lengths are as short as possible. Bio-Rad provides precut and
1/4-28 fitted, labeled tubing in the Fittings Kit. The illustration below shows where the tubing is designed to
be connected. Discussion of how to create your own tubing can be found in the following section.
Use 1/16” (1.6 mm) OD PEEK tubing and 1/4-28 fittings. Use orange PEEK 0.020” (0.51 mm) ID tubing for
the F10 pumps, and green PEEK 0.030” (0.76 mm) ID tubing for the F40 pumps. Premade tubing kits are
provided with your Fittings kit.
Figure 4-1. System Plumbing using Tubing Kit from Fittings Kit
COLUMN
AVR7-3 SAMPLE
INJECT VALVE
UV DETECTOR
CONDUCTIVITY
MONITOR
TO FRACTION
COLLECTOR
A
B
INLET A/B
INLET A/B
PUMP
#1
#5
#6
#4
#3
PUMP
MIXER
QUADTEC DETECTOR
CONDUCTIVITY
MONITOR
#3
WASTE
WASTE
#2
COLUMN
COLUMN
MIXER
LOOP
AVR7-3 INJECT VALVE
7
1
2
3
4
5
INJECT
PORT
6
#2
SYSTEM PLUMBING
SYSTEM INSTALLATION AND SETUP
4-1
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System