Bio-Rad BioLogic Signal Import Module (SIM) User Manual
Page 32

2.4.2
Maximizer Mixer
The Maximizer mixer provides the larger capacity required when the Maximizer is used. The Maximizer
mixer may be used with or without a mixer barrel: the mixer body, mixer top, and mixer barrel are provided
with the system; an optional mixer barrel extender is available. The mixer barrels fit between the mixer body
and the mixer top and are used for higher flow rates. Refer to the following table to determine the
appropriate mixer volume.
Table 2-11.
Mixer Barrels and Mixer Capacity for the Maximizer Mixer
Flow Rate
Barrel Extension
Capacity
Assembly Screws
0.5 to 10 ml/min
none
750 µl
1/4
-20 x
1/2
” (1.3 cm)
10 to 40 ml/min
5 ml Barrel Extender
5 ml
1/4
-20 x 1
1/2
” (3.8 cm)
40 to 80 ml/min
12 ml Barrel Extender
12 ml
1/4
-20 x 3
1/4
” (8.2 cm)
The procedure for changing the mixer capacity is provided in the following section.
Figure 2-3. Maximizer Mixer and Mixer Barrel Extender
MIXER
TOP
MIXER
BODY
12 ml MIXER
BARREL EXTENDER
OUTLET PORT
INLET PORT
5 ml MIXER
BARREL EXTENDER
MAXIMIZER MIXER
DESCRIPTION OF BIOLOGIC DUOFLOW SYSTEM
SYSTEM OVERVIEW
2-19
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System