Refer to Figure 11-6 for re-assembling the mixer. Make sure that the magnetic stir bar lies flat and
that the O-ring groove in either the mixer barrel or the mixer barrel extender faces up.
Place an O-ring in each O-ring groove. If you are using only the mixer body, then only one O-ring is
required. If either the mixer barrel or the mixer barrel extender is used, then two O-rings are
required.
Figure 11-6. Mixer Assembly
SCREWS
MIXER
BODY
O-RING
750
µ
l MIXER
BARREL EXTENDER
OUTLET PORT
O-RING
INLET PORT
MIXER
TOP
MAGNETIC
STIR BAR
O-RING GROOVE
O-RING GROOVE
ASSEMBLY OF THE MX-1 MIXER
MIXER
TOP
SCREWS
MIXER
BODY
12 ml MIXER
BARREL
OUTLET PORT
5 ml MIXER
BARREL EXTENDER
INLET PORT
MAGNETIC
STIR BAR
O-RING
O-RING GROOVE
O-RING GROOVE
ASSEMBLY OF THE MAXIMIZER MIXER
MAINTENANCE
MAINTENANCE AND TROUBLESHOOTING
11-9
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System