USB BITBUS COMMUNICATOR CABLE CONNECTIONS
The USB Bitbus Communicator allows communication between the Controller and the rest of the DuoFlow
system. To connect the USB Bitbus Communicator:
1.
Connect System Cable 31 from the USB Bitbus Communicator to the computer USB connector.
Refer to the illustration below.
2.
If a Signal Import Module (SIM) is to be used with the system (see lower half of Figure 3-3).
a. Connect the USB Bitbus Communicator power adapter (catalog number 760-2034).
b. Set the Power Select switch to External.
Figure 3-3. USB Bitbus Communicator Cabling
REAR VIEW
FRONT VIEW
REAR VIEW
USB BITBUS
COMMUNICATOR
SIGNAL IMPORT MODULE
USB BUS
REAR VIEW
EXT
PWR
SELECT
INT
FRONT VIEW
DELL CONTROLLER
REAR VIEW
DELL CONTROLLER
USB BUS
REAR VIEW
EXT
PWR
SELECT
INT
CONNECTION TO THE
FOLLOWING DEVICES
(LISTED IN SEQUENCE):
MAXIMIZER (IF AVAILABLE)
WORKSTATION
INSTRUMENT BUS
INSTRUMENT BUS
UNIVERSAL
AC/DC INLINE ADAPTOR
WALL AC POWER
CONNECTION TO THE
FOLLOWING DEVICES
(LISTED IN SEQUENCE):
SIM (IF AVAILABLE) OR
MAXIMIZER (IF AVAILABLE)
WORKSTATION
SYSTEM
CABLE 31
SYSTEM
CABLE 31
SYSTEM CABLE
17, 18, 19, 21, OR 30
SYSTEM CABLE
17, 18, 19, 21 OR 30
USB BITBUS
COMMUNICATOR
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System