Bio-Rad BioLogic Signal Import Module (SIM) User Manual
Page 71

DESCRIPTION OF SYSTEM COMPONENTS
SYSTEM OVERVIEW
2-58
2.10.11 Column Fittings
Table 2-15.
Columns and Column Fittings
Description
The Econo-Pac cartridge has one male and one
female luer-lock fitting. To connect it to the
BioLogic system use 1/4-28 to male and 1/4-28 to
female luer adapters, provided in the kit shown to
the right. One set is included in the Fittings kit.
Econo-Columns have two male luer-lock fittings.
To connect to the system, use female luer to 1/4-
28 adapters.
If a flow adapter with flexible tubing is to be used,
attach the barbed-to-male luer fitting (included in
this kit), then connect the female luer to 1/4-28
adapter.
If the column adapter uses
PTFE tubing, attach
the small length of 1/16” (1.6 mm) ID Tygon®
tubing and insert the barbed male luer fitting into
open end. Five cm of Tygon tubing is included in
this kit. Connect as above.
Pharmacia’s FPLC columns use M-6 fittings. To
connect to the system, use 1/4-28 to M6 adapter
union. Two of these are supplied with the Fittings
kit.
Fittings Required
Catalog # 732-0113
Econo-Pac
®
Cartridge to
Fittings Kit
Catalog # 750-0565
Econo-Column
®
to
BioLogic System Fittings Kit
This kit includes all the parts
needed to connect Econo-
Columns and flow adapters
to the BioLogic DuoFlow
system.
Catalog # 750-0561
Union, 1/4-28 to M6
Column Type
Econo-Pac®
Cartridge
Econo-Column®/
Low Pressure
Columns and flow
adapters
FPLC® Column
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System