2.8.3 EGP Econo Gradient Pump
The Econo Gradient Pump (EGP) is a two-channel, bi-directional, variable speed peristaltic pump for low-
pressure chromatography and general laboratory use. The EGP has the following features:
•
The EGP can be programmed to run a gradient.
•
The EGP controls a splitter valve for stream splitting.
•
The EGP may be used with most flexible tubing having an inner diameter less than or equal to 3.2
mm (1/8”), and a wall thickness of 1.0 mm or less, including Silicone, Tygon, and PharMed.
•
Flow rates are 0.01-20 ml/min per channel, depending upon tubing ID.
•
It has start/stop control of fraction collectors and chart recorders.
•
It is coldroom compatible.
Figure 2-25. EGP Econo Gradient Pump
Consult the Econo gradient pump user manual for additional information.
ECONO GRADIENT PUMP
Remote
Setup
Manual
Program
Mode
Step
Cancel
Next
Run
Purge
Reverse
LCD Display
Soft Keys
Arrow Keys
Platen
DESCRIPTION OF SYSTEM COMPONENTS
SYSTEM OVERVIEW
2-43
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System