11.7 MAXIMIZER VALVES
Over time, the Maximizer valves may become worn and require replacement. Indications of valve damage
include valve leaks and valve switching problems. To replace a valve, you’ll need a standard 9/64” hex key;
one has been provided for this purpose.
Safety Note: Before removing a valve, be sure to flush any hazardous material from the valve. Wear
protective clothing, as appropriate.
To remove the valve, follow the procedure below and refer to Figure 11-9:
1.
Use the 9/64” hex key to remove the two screws that attach the valve to the Maximizer. Set the
screws aside.
2.
Lower the front of the valve to disengage the outlet tubing from the top of the valve.
3.
Slowly pull away the valve to expose the wiring behind the valve.
4.
Using your fingers, press the connector latch to separate the valve wiring from the Maximizer wiring.
Figure 11-9. Replacing a Maximizer Valve
To attach the new valve, follow the procedure below:
1.
Connect the valve wiring to the Maximizer wiring.
2.
Place the valve against the Maximizer, but do not reattach the screws at this time. Rotate the valve
slightly downward until you can engage the valve with the output tubing to the Workstation.
Reattach the output tubing using its fittings.
3.
Reattach the valve mounting screws.
B1
WATER
B2
SAL
T
VAL
VE B
10
10
B1
WA
TER
B2
SAL
T
VAL
VE B
REMOVE MAXIMIZER VALVE
DISCONNECT MAXIMIZER VALVE
MAINTENANCE
MAINTENANCE AND TROUBLESHOOTING
11-13
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System