Bio-Rad BioLogic Signal Import Module (SIM) User Manual
Page 187

MULTIPLE COLUMNS
ADVANCED SYSTEM APPLICATIONS
9-6
Writing the Protocol
The powerful queuing feature of the BioLogic software makes writing multi-dimensional chromatography
experiments simple. Multi-dimensional experiment can be written as a series of individual one-dimensional
chromatography methods that are then ran sequentially as part of a queued run. Once written, each one-
dimensional method can be combined with other methods to make the desired experiment. For example, a
single affinity protocol could be could be combined with either a desalting method or size exclusion method
to create one of the two-dimensional experiment shown in Table 1. On the other hand an affinity method
could be combined with both a desalting method and an ion exchange method to create the three-
dimensional experiment shown in Table 1.
When writing a multi-dimensional method, care should be taken to prevent reintroduction of impurities
removed in previous dimensions. Cleaning and equilibration steps should be included at the start of each
dimension to wash any impurities from the tubing and to fill the system with the appropriate buffers.
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System