Modes of operation system operation 7-27 – Bio-Rad BioLogic Signal Import Module (SIM) User Manual
Page 156

Scouting Wizard Step 3
This dialog is used to set the number of runs to be performed, name each run and set the parameters for
the scouted steps.
•
Number of runs: Sets the number of runs to be performed as part of the scout experiment.
•
Auto-name: When selected, this feature automatically generates run names based on the current
Run 1 run name. Refresh updates the run names.
•
Auto-increment: Automatically generates scout run parameters by incrementing the Run 1 scout
parameter by the Increment Value. The scout Increment Value can be positive or negative. Note, you
should create your protocol with the appropriate starting value before defining the scout.
•
Finish: Saves the scout setup. Note, the protocol cannot be edited once a scout has been defined
unless the scout step is deleted.
Scouting Wizard - Step 3
Next > >
Finish
< < Back
Cancel
Refresh
Auto-naming
5
Number of Runs:
Gradient initia
Run 1
Scout %B (Gradient, Intial %B) 0.00
Run Name
%
0.00
5
Increment Value:
Auto-increment
Run 2
Scout %B (Gradient, Intial %B) 5.00
5.00
Run 3
Scout %B (Gradient, Intial %B) 10.00
10.00
Run 4
Scout %B (Gradient, Intial %B) 15.00
15.00
Run 5
Scout %B (Gradient, Intial %B) 20.00
20.00
MODES OF OPERATION
SYSTEM OPERATION
7-27
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System