Appendix a. specifications – Bio-Rad BioLogic Signal Import Module (SIM) User Manual
Page 220

SPECIFICATIONS
APPENDIX A
A-1
APPENDIX A.
SPECIFICATIONS
System Power Requirements: BioLogic DuoFlow Controller, Workstation, and Maximizer
•
100 V ~ 5.3 A
•
230 V ~ 3.5 A
•
50 - 60 Hz
DuoFlow Dell Controller
•
Pentium IV
®
Processor, 20 GB hard drive, 256 MB RAM
•
SoundNIC and video 4MB graphics performance accelerator
•
Dell M781 multimedia 16” monitor
•
3.5” floppy drive; CD-RW CD-ROM drive
•
Windows
®
2000 operating system
•
Microsoft
®
PS/2 IntelliMouse
™
•
DuoFlow software, version 5.0 or higher
•
Networkable for printer and file access
Maximizer
•
Size: 18.5” x 14.25” x 4.33”; 47 x 36 x 11 cm (L x W x H)
•
Weight: 11 lbs; 4.8 kg
•
Environmental: Cold room compatible
•
Solvent compatibility: 1M sodium hydroxide (NaOH), hydrochloric acid (HCl), and organic acids;
0.1M sulfuric acid; 7M urea and guanide hydrochloric acid; 100% methanol (MeOH), isopropanol
(IPA), and acetonitrile; 0.1% triflouracetic acid (TFA); 1% detergents (including sodium dodecyl
sulfate (SDS)) and TritonX-100
Workstation pump
•
Dual piston, positive displacement solvent pumps, with interchangeable solvent delivery pumpheads
(F10 and F40) made of bio-compatible PEEK
•
F10 pumpheads: Without Maximizer: 3,500 psi (233 bar, 23 MPa) from 0.01 to 10.0 ml/min in 0.01
ml/min increments. With Maximizer: 3,500 psi (233 bar, 23 MPa) from 0.5 to 20.0 ml/min in 0.02
ml/min increments.
Flow accuracy: 0.01 to 10.0 ml/min: 2%
Flow reproducibility: 0.01 rsd
•
F40 pumpheads: Without Maximizer: 1,000 psi (66 bar, 6.6 MPa) from 0.5 to 40.0 ml/min in 0.01
ml/min increments. With Maximizer: 1,000 psi (66 bar, 6.6 MPa) from 1.0 to 80.0 ml/min in 0.02
ml/min increments
Flow accuracy: 0.5 to 40.0 ml/min: 2%
•
Compositional accuracy: ±1.0%
•
High pressure dynamic mixing. MX-1 mixer volume without barrel: 263 µl; mixer without mixer barrel
extender: 750 µl; mixer with mixer barrel extender: 2.0 ml. Maximizer mixer volume without barrel:
750 µl; mixer with mixer barrel extender: 5.75 ml; mixer with expansion barrel: 12.75 ml.
•
Built-in pressure transducer
•
Gradient by software control: 1% increments
- BioLogic DuoFlow Pathfinder 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Pathfinder 20 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 80 System DuoFlow™ Chromatography System BioLogic DuoFlow Maximizer 20 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 40 System DuoFlow™ Chromatography System BioLogic DuoFlow QuadTec 10 System DuoFlow™ Chromatography System BioLogic DuoFlow 40 System Chromatography System BioLogic DuoFlow 10 System Chromatography System