MEASUREMENT

75

-Pulses

Measures the number of
negative pulses.

+Edges

Measures the number of
positive edges.

-Edges

Measures the number of
negative edges.

Delay
Measurement

FRR

Time between:

Source 1 first rising edge and
Source 2 first rising edge.

FRF

Time between:
Source 1 first rising edge and
Source 2 first falling edge.

FFR

Time between:
Source 1 first falling edge
and Source 2 first rising
edge.

FFF

Time between:
Source 1 first falling edge
and Source 2 first falling
edge.

LRR

Time between:
Source 1 first rising edge and
Source 2 last rising edge.

LRF

Time between:
Source 1 first rising edge and
Source 2 last falling edge.

LFR

Time between:
Source 1 first falling edge
and Source 2 last rising edge.

LFF

Time between:
Source 1 first falling edge
and Source 2 last falling
edge.