XY Mode

Cursors can be used with XY mode.
See the Cursor chapter for details.

Page 85

Set the Sampling Mode

Background

The GDS-2000A has two types of sampling modes:
ET (Equivalent Time) and Sin(x)/x interpolation.
Equivalent time sampling is able to achieve a
sample rate of 100GSa/s when sampling periodic
waveforms. Sin(x)/x interpolation uses a sinc
interpolation formula to reconstruct a continuous
signal between sampled points.

Sin(x)/x

One sample of data is used to
reconstruct a single waveform.
Sin(x)/x sampling should be used
when the time base is relatively
slow or if single shot events need to
be captured.

Equivalent-
time sampling

Sampled data is accumulated a
number of times to reconstruct a
single waveform. This increases the
sampling rate, but can only be used
for repetitive signals. This mode is
usually used when the time base is
too fast for real-time sampling.

Panel Operation

1. Press the Acquire key.

Acquire

2. Press the ET/sin(x)/x key on the

bottom menu to toggle between
equivalent time sampling (ET) and
sin(x)/x interpolation.