beautypg.com

GW Instek GDS-2000A series User Manual User Manual

Page 180

background image

GDS-2000A Series User Manual

180

Range

Voltage division range

Save Boundary
Template

4. Press Save Operation to save the

tolerance boundaries.

5. The Maximum Position tolerance

will be saved to reference
waveform R1, and the Minimum
Position tolerance to R2.

6. Press Go Back to return to the

previous menu.

Note

It may be necessary to save a reference waveform to
R1 and R2 before the tolerance for Maximum and
Minimum Position can be created. This can be done
by first creating and saving the Auto Tolerance
boundary.

Start Go-NoGo

Press Enable to start the Go-NoGo test.
The Enable button will change to
Disable. Pressing Disable will stop the
Go-NoGo test and toggle the button
back to Enable.

If the Violating setting was set to Stop
or Stop Beep, press Enable to restart
the test after it has stopped.

Results

When Go-NoGo is running, the pass/fail ratio is
displayed in the bottom left-hand corner. The first
digit represents the number of failed tests, and the
right hand digit represents the number of tests.